A blood oxygenator for a 500 g patient: 8 mL to fill, no pump, driven by the infant’s own heart.

In development. What you read here are design targets, not test results.

The smallest oxygenator a hospital can buy holds 40 mL of blood. An infant born at 500 g has about 45 mL in total. Novanatomy is designing one that fills with 8.

Who it is for. Infants born at 22 to 25 weeks, weighing 400 to 1,000 g. Their lungs are not ready for air, and ventilating them can cause lasting injury. Support outside the body could spare the lungs, but no circuit on the market is sized for a patient this small.

What we are making. A single-use hollow-fibre cartridge with a matched cannula set. About 0.1 m² of membrane, 8 mL of fill or less, and resistance low enough that the infant’s own heart moves the blood. Nothing else sits in the blood path.

Where it stands. Requirements and sizing are drafted and prototypes come next. Every figure on this page is a design target, not a measured result.

Who we want to hear from. Neonatologists, fetal surgeons, ECMO coordinators and perfusionists who can tell us where these targets are wrong. Laboratories running preterm lamb or piglet models. Engineers who have built hollow-fibre devices.

500 g

400 g1,000 g

How much blood fills the machine?

A circuit has to be filled before it can run, and whatever fills it mixes with the patient’s blood. When the fill is nearly as large as the patient’s whole blood volume, the circuit must be primed with donor blood.

  • The infant’s blood: about 45 mL
  • Today’s smallest neonatal oxygenator: 40 mL to fill
  • The Novanatomy target: 8 mL to fill

At 500 g, filling today’s smallest oxygenator takes 89% of the infant’s blood volume. The target cartridge takes 18%.

Areas are to scale. Blood volume is estimated at 90 mL per kg, a textbook approximation. The 40 mL figure is the published fill volume of the smallest commercial neonatal oxygenator, which has 0.38 m² of membrane.

Two arteries out,one vein back

The circuit copies the route blood already takes to the placenta. The cannulas use half of the pressure the heart can offer, so they are designed together with the cartridge.

  1. Leave

    Blood low in oxygen leaves through the two umbilical arteries. Only the infant’s heart pushes it: about 20 mmHg of driving pressure.

  2. Cross

    It passes once through a short, wide bundle of hollow fibres. Oxygen moves in and carbon dioxide out, for 5 mmHg or less.

  3. Return

    Oxygenated blood returns through the umbilical vein. No pump, heat exchanger, filter or reservoir stands in the way.

The design targets

Electron micrograph of a single hollow fibre cut across, showing its open bore and porous wall, with a 500 micrometre scale bar
One hollow fibre in cross-section. Gas runs through the bore; blood flows around the outside. A polysulfone fibre, shown for scale, not ours.

8 mL

to fill the cartridge, against 40 mL for today’s smallest.

28 days

of continuous support without a change-out. The floor is 14.

Pressure, not gas exchange, is the hard constraint. Membrane area can shrink fourfold, but resistance has to stay as low as a device four times the size.

ParameterMust meetAim for
Patient weight400 to 1,000 gDesign point 500 g
Blood flow40 to 300 mL/minStable down to 30 mL/min
Oxygen transfer8 mL/min at 150 mL/min of flow12 mL/min
Carbon dioxide removal8 mL/min12 mL/min
Cartridge fill8 mL or less5 mL or less
Whole circuit fill15 mL or less10 mL or less
Pressure drop5 mmHg or less at 100 mL/min3 mmHg or less
Continuous run14 days28 days, no change-out
Blood thinnersLow-dose heparinNone
Red-cell damageNo worse than today’s neonatal deviceHalf

Nothing but fibrein the blood path

Every part added to a circuit adds volume to fill and surface to clot on. The baseline is a fibre bundle in a clear housing, with a port at its high point to let the air out.

Talk to us

Nine tests beforeany patient

Each prototype climbs the same ladder, cheapest test first, and only a design that passes one rung moves to the next. This is the plan, not a record. Results will be published here as they exist.

StageTestMethodPass
BenchPressure and flowGlycerol and water at blood viscosity, 20 to 400 mL/min5 mmHg or less at 100 mL/min
Fill and de-airingFill by weight; time to bubble-free8 mL or less, under 2 minutes
BloodGas transferOvine or bovine blood at standard and fetal conditions8 mL of oxygen a minute at 150 mL/min
Red-cell damageSix-hour loop beside today’s neonatal deviceNo higher than the control
ClottingLow-heparin and no-heparin loops, each coatingNo loss of flow over 6 hours
EndurancePumpless behaviourPulsing mock circulation at 20 to 25 mmHg150 mL/kg/min for a 500 g model
Durability30 days of continuous runningNo leak, under 20% loss of transfer
AnimalAcutePreterm lamb or piglet, 24 to 72 hours, with a partner laboratoryStable flow and blood gases
Chronic14 days, with a partner laboratory14 days of stable pumpless support

We want to hear from the people who would use this cartridge, test it, or tell us where it is wrong.

Line drawing of an early infant incubator in section, with an infant lying above a row of hot-water bottles
Tarnier’s incubator, from Budin’s The Nursling, 1907.

Run a NICU, an ECMO programme or a fetal lab?

Tell us where you work and which target you would change first.